Open original FDA record ↗Recall record
- FDA identifier
- Z-0521-2025
- Company in FDA record
- MicroSurgical Technology Inc
- FDA classification date
- Nov 22, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to pin coming off of handles.
Source facts
- Action
- On 09/24/2024, the firm sent via express mail a customer letter informing customers that there is a potential issue pertaining to the pins on the handle of the DFH-1028, Handle Assembly Lot Number 160458. The locking pins that actuate the head may be defective and could potentially dislodge without warning. As a precautionary measure, the firm is replacing the handle associated with the affected lot number 160458 at no cost to customers. Customers are being instructed to return the affected products using the supplied return shipping label. Firm's contact information if questions or further assistance is needed: David Wang Sr. VP of Ops QA/RA D: 206-402-1146 8415 154th Avenue NE | Redmond, WA 98052 www.microsurgical.com dwang@microsurgical.com
- Initiated
- 2024-09-17
- Root cause
- Process control
- Distribution
- U.S.: CT, GA, MN, MT and TX. O.U.S.: Netherlands 24 total devices = 6 devices distributed in the United States and 18 devices distributed in Germany.
- Recall number
- Z-0521-2025
- Recall status
- Open, Classified
- CDRH record ID
- 210638
- FDA classified
- 2024-11-22
- Recall event number
- 95572