Open original FDA record ↗Recall record
- FDA identifier
- Z-0522-2013
- Company in FDA record
- Roche Diagnostics Operations, Inc.
- FDA classification date
- Dec 13, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
It has been determined that cobas b 123 Fluid Pack COOX cuvettes for lot numbers 21426121 and 21426171, exhibit a significant increase in the out-of-box failure rate caused by a change in the thickness in the cuvette layer. Internal investigations found the altered thickness of the cuvette was due to production equipment settings. Cuvettes were produced within specifications, but within the upper
Source facts
- Action
- Roche sent an Urgent Medical Device Removal (UMDR12-268) notification and faxback form on November 14, 2012, to all affected consignees. Customers were requested to discontinue the use of cobas b 123 Fluid Pack COOX, 200 lots 21426121 and21426171 and discard the product per local regulations. Customers were instructed to use the enclosed faxback form to order replacement product at 1-800-722-7222. For questions customers were instructed to contact Roche Diagnostics Technical Support at 1-800-526-2272. For questions regarding this recall call 317-521-3966.
- Initiated
- 2012-11-14
- Root cause
- Software design (manufacturing process)
- Terminated
- 2013-08-07
- Distribution
- Nationwide Distribution including NE and AZ.
- Recall number
- Z-0522-2013
- Recall status
- Terminated
- CDRH record ID
- 114521
- FDA classified
- 2012-12-13
- Recall event number
- 63680