The previous software application version (1.1.300) is missing a decimal separator (a comma) for parameter range guidance values displayed on some of the programming screens: Catheter, Reservoir, Infusion, Bolus, myPTM, and Alarm.
Source facts
Action
Medtronic issued a SAFETY NOTIFICATION to its consignees on 07/13/2020 by letter. The notice explained the problem with the device and requested the software be upgraded to version 1.1.342.
Initiated
2020-07-13
Root cause
Software Design Change
Distribution
Foreign Distribution
Recall number
Z-0522-2024
Recall status
Completed
CDRH record ID
204258
FDA classified
2023-12-09
Recall event number
93401