Skip to content

Public FDA record

Z-0522-2024: Medtronic SynchroMed, Model A10

Recall record Z-0522-2024 for Medtronic Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0522-2024
Company in FDA record
Medtronic Inc.
FDA classification date
Dec 9, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The previous software application version (1.1.300) is missing a decimal separator (a comma) for parameter range guidance values displayed on some of the programming screens: Catheter, Reservoir, Infusion, Bolus, myPTM, and Alarm.

Source facts

Action
Medtronic issued a SAFETY NOTIFICATION to its consignees on 07/13/2020 by letter. The notice explained the problem with the device and requested the software be upgraded to version 1.1.342.
Initiated
2020-07-13
Root cause
Software Design Change
Distribution
Foreign Distribution
Recall number
Z-0522-2024
Recall status
Completed
CDRH record ID
204258
FDA classified
2023-12-09
Recall event number
93401

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗