Z-0522-2025: Kreatech PapV-16, Flu labeled ISH probe, to detect presence of HPV-16 DNA inserts in human tissue, Model

Recall record Z-0522-2025 for Kreatech Biotechnology Bv. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0522-2025
Company in FDA record
Kreatech Biotechnology Bv
FDA classification date
Nov 22, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The DNA fragments for this specific lot number are too large, resulting in potential background around the cells due to suboptimal hybridization. An absence of signal could also be observed as a result of the suboptimal hybridization of the larger DNA fragments. The user may interpret potential background as a false positive result and an absence of signal as a false negative result, if proper positive and negative controls are not included.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL notification letter dated 10/31/24 was sent to customers. Actions to Be Taken by the Customer/User 1. Do not use or continue to use the batch listed above. You will receive a replacement product as soon as possible. 2. Please dispose of any unused products according to your organization s disposal policies. 3. Review and follow up on any results obtained using the above-mentioned product batch. 4. Please read and retain this communication. Please confirm receipt of this letter as soon as possible by signing and returning the below Acknowledgement Form to: QA.Amsterdam@leicabiosystems.com. Transmission of This Notification Letter Kindly please pass this notification to the users of this product and to all those within your organization who need to be aware of this issue. If you have any questions about this notification letter, please contact your local Leica Biosystems representative or contact QA.Amsterdam@leicabiosystems.com.
Initiated
2024-11-01
Root cause
Under Investigation by firm
Distribution
US States: FL, NY
Recall number
Z-0522-2025
Recall status
Open, Classified
CDRH record ID
210918
FDA classified
2024-11-22
Recall event number
95668

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗