Potential for modules to reset, alarms returning to default settings, invasive pressure losing its labels and zero calibration and Sp02 alarm limits being frozen. This can result in delay in treatment.
Source facts
Action
On 1/3/07 the firm issued a letter to their customers advising of the issues and provided guidance to the consignee to enable the Check Setup feature that will provide an audible prompt whenever the monitor resets. The letter advises that consignees will be contacted to schedule an update, once new software has been validated.
Initiated
2007-01-03
Root cause
Other
Terminated
2011-02-15
Distribution
Nationwide and worldwide.
Recall number
Z-0523-2007
Recall status
Terminated
CDRH record ID
49866
FDA classified
2007-02-27
Recall event number
37077