Z-0523-2024: ARTIS icono (ceiling configuration), Fluoroscopic x-ray system; Model no. 11328100

Recall record Z-0523-2024 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0523-2024
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Dec 9, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential hardware issue for ARTIS icono ceiling system: collision sensor may falsely signal non-existing collisions for up to half an hour when the system is switched off for a longer time (e.g., overnight) and switched on. During this time, system movement is only possible with the "override" function with very slow speed and without further collision protection. This may lead to temporary unavailability of the system for diagnostic and therapeutic usage.

Source facts

Action
On October 17, 2023, the firm issued an URGENT MEDICAL DEVICE CORRECTION letter to affected customers. According to the firm's letter, as the unintended behavior is caused by thermal startup effects, its occurrence can be avoided by only switching off the system for a short time. The firm suggests keeping the system running (except during proposed weekly reboot) if availability for treatment of emergency patients is necessary. Please plan for up to half an hour warm-up time after switching on the system. The firm will contact affected customers to modify the mounting of the tube cover to mitigate the probability of occurrence of the issue. You may contact the firm's service organization for an earlier appointment at 1-800-888-7436.
Initiated
2023-10-17
Root cause
Component design/selection
Distribution
US Nationwide - Worldwide Distribution: Domestic distribution to FL GA IA IL IN MI MN MO MT PA TX. Foreign distribution to Australia Canada Switzerland China Germany Denmark Egypt Spain Finland United Kingdom Israel Netherlands Sweden South Africa
Recall number
Z-0523-2024
Recall status
Open, Classified
CDRH record ID
204030
FDA classified
2023-12-09
Recall event number
93338

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗