Z-0523-2025: Brand Name: Z-800 Infusion System Product Name: Z-800WF Model/Catalog Number: Z-800WF. The Z-800 Infusion

Recall record Z-0523-2025 for Zyno Medical LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0523-2025
Company in FDA record
Zyno Medical LLC
FDA classification date
Nov 22, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

For the Z-800WF pumps, the WiFi software was not compatible with the pump software that had been loaded in accordance with ongoing correction Z-1183-2024.

Source facts

Action
On September 30, 2024 MEDICAL DEVICE NOTIFICATION letters were sent to customers. Mitigation: Zyno Medical is issuing this customer notification, advising customers that the expected User ID may be missing from completed therapies in the EMR Patient Record. Zyno Medical is planning to correct this issue through an update to the software. Recommended Actions: The user can locate the Patient Order in the Patient Record in the EMR and manually update the record to reflect the administrating clinician s User ID until the Z-800WF pump software version is updated. Follow the guidelines of your facility for performing manual updates in the EMR. Reporting Adverse Events: We strongly encourage healthcare providers and patients to promptly report any adverse events associated with the use of the Zyno Medical Z-800, Z-800F, Z-800W, and/or Z-800WF Infusion Pump. Please send your reports to Zyno Medical via email to feedback@intuvie.com. Additionally, we urge you to report any negative side effects related to the Zyno Medical Z-800, Z-800F, Z-800W, and/or Z-800WF Infusion Pumps to the FDA. You can submit your report online at www.fda.gov/medwatch or by calling 1-800-FDA-1088. Your cooperation in reporting these events is vital for continuous safety monitoring of medical devices. For information about the Zyno Medical Z-800, Z-800F, Z-800W, and/or Z-800WF Infusion Pumps, including benefits and risks, refer to the Zyno Medical Z-800 Infusion Pump Instructions for Use (Z-800 800-IFU-6651, Z-800F 800-IFU-2602, Z-800W 800W-IFU-3614, and Z-800WF 800WF-IFU-4602) for complete description of the benefits and risks of using the Zyno Medical Z-800 family of pumps. Zyno Medical is committed to offering products which meet the highest quality standards and to continuous improvement in the interest of patient safety. We regret any inconvenience that this action may cause, but believe transparent communication will ensure you have timely, relevant information for managing your patients. If you
Initiated
2024-09-30
Root cause
Software design (manufacturing process)
Distribution
Domestic: AL, FL, LA, MI, NE, NJ, OH, TX.
Recall number
Z-0523-2025
Recall status
Open, Classified
CDRH record ID
210846
FDA classified
2024-11-22
Recall event number
95635

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗