On 10/27/2023, the firm sent via email an "URGENT: MEDICAL DEVICE RECALL" Letter to customer informing them that the firm was made aware of a mistake on the label of the HPS tibial liner #6H. 10mm.
Customers are instructed to:
Check their stock to locate and quarantine the affected devices. Devices must be sent back
to the Lima local reference within 15 days together with a hard copy of the attached
Acknowledgement and Receipt Form at the following address:
Lima U.S.A., Inc. | 2001 NE Green Oaks Blvd. Ste.100 | Arlington, Texas 76006 | USA
For questions or further assistance, email medicalcomplaints@limacorporate.com.
Initiated
2023-10-27
Root cause
Labeling Change Control
Distribution
U.S Distribution: AZ, CA, IL, LA, and NY
Recall number
Z-0524-2024
Recall status
Open, Classified
CDRH record ID
203991
FDA classified
2023-12-09
Recall event number
93329