a. A small number of Fusion Pro 24 units were released with an incorrectly placed Interlock device. As a result, the system can be operated with the front-loading door in an open position.
b. Fails to comply with Class 2 designation, indicated in Part 6 of Laser Product Report. An incorrectly placed Interlock device can allow Class 4 laser radiation fields during operation.
Source facts
Action
The measures to be taken to repair such defect or to bring the product into
compliance with the Federal standard.
This Recall provides corrections by repair to a defective safety interlock location by a kit with comprehensive instructions, offered at no cost to purchasers
Initiated
2024-08-08
Root cause
Radiation Control for Health and Safety Act
Distribution
US Nationwide Distribution
Recall number
Z-0524-2025
Recall status
Open, Classified
CDRH record ID
211119
FDA classified
2024-12-03
Recall event number
95757