Skip to content

Public FDA record

Z-0525-06: Bard Composix Kugel Mesh X-Large Patch Oval with ePTFE, 10.8'' x 13.7'' (25.4cm x 33cm) Code: 0010207

Recall record Z-0525-06 for Davol, Inc., Sub. C. R. Bard, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0525-06
Company in FDA record
Davol, Inc., Sub. C. R. Bard, Inc.
FDA classification date
Feb 22, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Memory recoil ring could break and potentially lead to bowel perforation and or chronic enteric fistula

Source facts

Action
Davol Inc. notified US Customers by letter on 12/27/05 via Federal Express. Bard Regulatory International was notified 12/22/05 to recall to accounts outside the US. Accounts are requested to return recalled product. A second letter issued dated January 5, 2006 to include additional product codes not identified in the first letter, specifically to the accounts that received those lots. On 3/24 / 06, Davol issued a third Recall Notification for the extended recall to Hospitals and Distributors and a Dear Doctor Letter to Chief of Surgery.
Initiated
2005-12-22
Root cause
Other
Distribution
Nationwide Foreign: Australia, Canada, Japan,
Recall number
Z-0525-06
Recall status
Completed
CDRH record ID
43768
FDA classified
2006-02-22
Recall event number
34393

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗