Skip to content

Public FDA record

Z-0525-2013: Praxair Vantage Grab n Go - Gas Cylinder pressure regulator Catalog Number: OX-M-AEGNGVNTG Usage: Regul

Recall record Z-0525-2013 for Praxair Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0525-2013
Company in FDA record
Praxair Inc.
FDA classification date
Dec 21, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class I

Reason stated in the source

Isolated incidents of ignition inside Grab n' Go cylinders that had been knocked over or otherwise subject to significant physical impact.

Source facts

Action
The firm, PRAXAIR, sent a "MEDICAL DEVICE CORRECTION" letter dated November 26, 2012 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to follow the Grab'n Go operating instructions provided in the letter which include: storing the Grab'n Go systems indoors in a safe area; never store a system in an elevated location; handle systems safely and avoid jarring or dropping the system and locate at the following web link: www.praxair.com/GNGinstructions. The customers were also instructed to ensure that all of the Grab'n Go units that were sent by Praxair are sent to Praxair for servicing and refilling after use, in accordance with their normal process. If you have any questions, contact your sales representative at 630-320-4452 or email: Thelma_Brantley@Praxair.com.
Initiated
2012-11-16
Root cause
Device Design
Terminated
2014-03-24
Distribution
Worldwide distribution: USA (nationwide) and countries of: Brazil, Canada, China, Columbia, Costa Rica, India, Paraguay and Uruguay.
Recall number
Z-0525-2013
Recall status
Terminated
CDRH record ID
114654
FDA classified
2012-12-21
Recall event number
63729

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗