Open original FDA record ↗Recall record
- FDA identifier
- Z-0525-2013
- Company in FDA record
- Praxair Inc.
- FDA classification date
- Dec 21, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class I
Reason stated in the source
Isolated incidents of ignition inside Grab n' Go cylinders that had been knocked over or otherwise subject to significant physical impact.
Source facts
- Action
- The firm, PRAXAIR, sent a "MEDICAL DEVICE CORRECTION" letter dated November 26, 2012 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to follow the Grab'n Go operating instructions provided in the letter which include: storing the Grab'n Go systems indoors in a safe area; never store a system in an elevated location; handle systems safely and avoid jarring or dropping the system and locate at the following web link: www.praxair.com/GNGinstructions. The customers were also instructed to ensure that all of the Grab'n Go units that were sent by Praxair are sent to Praxair for servicing and refilling after use, in accordance with their normal process. If you have any questions, contact your sales representative at 630-320-4452 or email: Thelma_Brantley@Praxair.com.
- Initiated
- 2012-11-16
- Root cause
- Device Design
- Terminated
- 2014-03-24
- Distribution
- Worldwide distribution: USA (nationwide) and countries of: Brazil, Canada, China, Columbia, Costa Rica, India, Paraguay and Uruguay.
- Recall number
- Z-0525-2013
- Recall status
- Terminated
- CDRH record ID
- 114654
- FDA classified
- 2012-12-21
- Recall event number
- 63729