Skip to content

Public FDA record

Z-0526-03: Philips BV300 Mobile X-Ray System

Recall record Z-0526-03 for Philips Medical Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0526-03
Company in FDA record
Philips Medical Systems
FDA classification date
Feb 1, 2003
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The C-bow support can bend and break allowing C-arm to fall from support

Source facts

Action
The firm issued a letter dated 9/27/02 to their customers. The letter advised of the problem and that a service representative will visit to inspect unit and correct.
Initiated
2002-09-27
Root cause
Other
Terminated
2005-01-12
Distribution
The firm distributed to hospitals, x-ray centers and medical centers throughout the U.S.
Recall number
Z-0526-03
Recall status
Terminated
CDRH record ID
25458
FDA classified
2003-02-01
Recall event number
25248

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗