Z-0526-2022: 1. Footrest ( use exclusively in combination with Intelli-C devices) 2. Shoulder support ( use exclusivel

Recall record Z-0526-2022 for NRT X-RAY A/S. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0526-2022
Company in FDA record
NRT X-RAY A/S
FDA record creation date
Jan 25, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Friction rubber block in the clamping mechanism was missing in the Footrest. It may slip down along the patient table while in use with a patient. This accessory is designed to use with Intelli-C, Right and Intelli-C, Left.

Source facts

Action
On 01/11/2022, NRT issued an Urgent Medical device correction, which includes the following action to be taken by the users. 1. Identified the affected accessories as described in the Urgent Medical Device Correction Letter. 2. Inspect the Footrest and Shoulder Support which use the same clamping mechanism and should be inspected in the same way. The clamping mechanism consists of two clamping profiles, one for each side of the tabletop. Both sides should be inspected. a. The friction rubber that must be inspected is placed inside the clamping profile. It is the thick rubber, which must be firmly in place. The rubber is glued on the profile, and you should look for the overall adhesion. If slightly pressed from each side and end, the rubber should not show signs of detachment from the profile. If the rubber is firmly in place, you may continue use the equipment. If the rubber is detaching along either side or end, do not use the accessory until the repair has been made. 3. Return the response card in page 4, with the results and date of the inspection and your signature, not later than February 15, 2022. to support@nrtxray.com. 4. Follow up will be made in regards to the repair in April 2022, which will be coordinated with US dealer Alpha Imaging in Willoughby, Ohio. For any question contact (300) 696-0004 or E-Mail at ksmith@alpha-imaging.com.
Initiated
2022-01-21
Root cause
Process control
Distribution
USA: IA, KY,LA, MA, NC,NY, OH, SC, WA, WV OUS: Switzerland, Czech Republic, Germany, Denmark, Netherlands, RSA, Sweden
Recall number
Z-0526-2022
Recall status
Open, Classified
CDRH record ID
191503
FDA record created
2022-01-25
Recall event number
89413

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗