Open original FDA record ↗Recall record
- FDA identifier
- Z-0527-2014
- Company in FDA record
- Karl Storz Endoscopy America Inc
- FDA classification date
- Dec 17, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
KARL STORZ Endoscopy America, Inc. is recalling the Flexible Broncho Fiberscope or a Flexible Intubation Fiberscope with a suction port because the material used in the suction channel portion was thermoplastic polyurethane (TPU), which is not listed in STERRAD systems claims for interior endoscope lumens.
Source facts
- Action
- The firm, KARL STORZ, sent a "A SAFETY ALERT-ATTENTION NEEDED" letter dated November 8, 2013 to all their customers on November 13, 2013. The letter describes the product, problem and actions to be taken.The letter informs the customers that the endoscopes incorporated a lumen material that is only approved for use as an external surface material and not included in the STERRAD NX and 100NX FDA-cleared claims for use in flexible endoscope lumens. The customers were instructed to follow the recommended processing methods included in the letter. Customers with questions about the letter are instructed to contact the undersigned and the Technical Services at (800) 421-0837, ext 5350.
- Initiated
- 2013-11-13
- Root cause
- Under Investigation by firm
- Terminated
- 2014-03-27
- Distribution
- Nationwide Distribution.
- Recall number
- Z-0527-2014
- Recall status
- Terminated
- CDRH record ID
- 123807
- FDA classified
- 2013-12-17
- Recall event number
- 66862