Open original FDA record ↗Recall record
- FDA identifier
- Z-0527-2019
- Company in FDA record
- Barco N.V.
- FDA record creation date
- Nov 30, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
If metal strips are removed and re-applied to another display, there is potential for the metal strips to be deformed resulting in a smaller contact surface with the front cover magnet strips. There is a risk that the protective front cover could come loose and may fall down.
Source facts
- Action
- Customers were notified via email on about 03/30/2017. Customers were alerted that a safety related update has been made to the installation manual. A safety notice has been shipped to distributors with the updated installation instructions manual and an acknowledgement letter. Distributors were asked to do the following: a) Inform all your customers who have purchased the Front Cover Kit by sending them the Safety Notice; b) Send the updated installation instructions to all your customers who have purchased the Front Cover Kit; c) Fill out the acknowledgment letter; attach and send in reply to this email.
- Initiated
- 2017-03-30
- Root cause
- Device Design
- Terminated
- 2020-05-04
- Distribution
- Worldwide - US Nationwide Distribution to CA, IL, GA, TX, and WI. In the countries of Belgium, Brazil, Canada, Czech Republic, Denmark, Finland, France, Germany, Italy, Japan, Poland, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Taiwan, Turkey.
- Recall number
- Z-0527-2019
- Recall status
- Terminated
- CDRH record ID
- 167949
- FDA record created
- 2018-11-30
- Recall event number
- 81239