Open original FDA record ↗Recall record
- FDA identifier
- Z-0527-2022
- Company in FDA record
- Velano Vascular
- FDA record creation date
- Jan 25, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Incomplete sealing of blood collection device unit packages may produce a breach in the sterile barrier enclosing the product. This could lead to microbial contamination of the device prior to use. The health effects range from no clinical effect up to bloodstream infection.
Source facts
- Action
- On 11/10/21, recall notices were sent to Medical Directors, Risk Managers, Medical Device Safety Officers, Nurse Managers. 1) Locate and destroy all product subject to the recall following your institution s process for destruction. 2) If the affected product was previously used in patient care and no signs of infection are apparent, no additional steps need to be taken. If the affected product was used in patient care and signs of infection are present, follow your institution s established procedures. 3) Share this recall notification with all users within your facility network of the product to ensure they are also aware of this recall. If you further distributed product, identify customers within your distribution network that purchased the affected product, as defined in this recall notification. Email an excel file listing of all customers to BDRC9@bd.com so the recalling firm may initiate customer notification. 4) Complete and return the Distributor Response Form. Customers require further assistance are encouraged to call/email: 1-844-823-5433/ productcomplaints@bd.com
- Initiated
- 2021-11-10
- Root cause
- Process control
- Terminated
- 2023-07-31
- Distribution
- US Nationwide distribution in the states of CT, CO, UT, PA, IL, WA, CA, NJ, MI, FL, NC.
- Recall number
- Z-0527-2022
- Recall status
- Terminated
- CDRH record ID
- 190721
- FDA record created
- 2022-01-25
- Recall event number
- 89185