Skip to content

Public FDA record

Z-0527-2022: PIVO Blood Collection Device 20G, REF: 202-0005

Recall record Z-0527-2022 for Velano Vascular. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0527-2022
Company in FDA record
Velano Vascular
FDA record creation date
Jan 25, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Incomplete sealing of blood collection device unit packages may produce a breach in the sterile barrier enclosing the product. This could lead to microbial contamination of the device prior to use. The health effects range from no clinical effect up to bloodstream infection.

Source facts

Action
On 11/10/21, recall notices were sent to Medical Directors, Risk Managers, Medical Device Safety Officers, Nurse Managers. 1) Locate and destroy all product subject to the recall following your institution s process for destruction. 2) If the affected product was previously used in patient care and no signs of infection are apparent, no additional steps need to be taken. If the affected product was used in patient care and signs of infection are present, follow your institution s established procedures. 3) Share this recall notification with all users within your facility network of the product to ensure they are also aware of this recall. If you further distributed product, identify customers within your distribution network that purchased the affected product, as defined in this recall notification. Email an excel file listing of all customers to BDRC9@bd.com so the recalling firm may initiate customer notification. 4) Complete and return the Distributor Response Form. Customers require further assistance are encouraged to call/email: 1-844-823-5433/ productcomplaints@bd.com
Initiated
2021-11-10
Root cause
Process control
Terminated
2023-07-31
Distribution
US Nationwide distribution in the states of CT, CO, UT, PA, IL, WA, CA, NJ, MI, FL, NC.
Recall number
Z-0527-2022
Recall status
Terminated
CDRH record ID
190721
FDA record created
2022-01-25
Recall event number
89185

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗