Open original FDA record ↗Recall record
- FDA identifier
- Z-0527-2025
- Company in FDA record
- Smiths Medical ASD, Inc.
- FDA classification date
- Nov 25, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.
Source facts
- Action
- Smiths Medical issued an URGENT MEDICAL DEVICE REMOVAL notice to its consignees on 10/28/2024, mailed using a traceable means. The notice explained the issue, potential risk, and requested the consignee locate and destroy all affected products, and distribute the notice to all affected parties. Distributors were directed to notify their customers. For questions: About the communication: smithsmedical7602@sedgwick.com, 1-800-913-3897; To report adverse events: globalcomplaints@icumed.com, 1-866-216-8806
- Initiated
- 2024-10-28
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide distribution.
- Recall number
- Z-0527-2025
- Recall status
- Open, Classified
- CDRH record ID
- 210757
- FDA classified
- 2024-11-25
- Recall event number
- 95605