Open original FDA record ↗Recall record
- FDA identifier
- Z-0528-05
- Company in FDA record
- Hu-Friedy Mfg Co, Inc.
- FDA classification date
- Feb 4, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Potential for tip breakage at the terminal shank due to improper heat-treatment at the foreign plant. Several reports of tip breakage were received by the own label distributor, and forwarded to the importer of record & specification setter of the dental explorers.
Source facts
- Action
- Hu-Friedy Mfg. Co., Inc. notified the Own Label Distributor of the recall on 1/7/05 by phone, and in writing on 1/10/05. Henry Schein, Inc. is having formal "URGENT: DEVICE RECALL" letters printed up which will be mailed to over 8100 of Schein's direct customers by First Class Mail on/about 1/21/05 to initiate the sub-recall of the dental explorers. Henry Schein will follow-up to non-responders by phone and/or fax.
- Initiated
- 2005-01-07
- Root cause
- Other
- Terminated
- 2012-05-24
- Distribution
- Dental practitioners, hospitals, dental clinics and distributors located nationwide, in Puerto Rico and internationally.
- Recall number
- Z-0528-05
- Recall status
- Terminated
- CDRH record ID
- 36764
- FDA classified
- 2005-02-04
- Recall event number
- 30882