Skip to content

Public FDA record

Z-0528-06: IMMULITE 2000 Total Testosterone Kit

Recall record Z-0528-06 for Diagnostic Products Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0528-06
Company in FDA record
Diagnostic Products Corp.
FDA classification date
Feb 16, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The reagent for this kit lot (L2KTW 2,6 126) was manufactured with a raw material that can carryover and interfere with DPC Kit IMMULITE 2000 Estradiol (L2KE2) causing a false elevation in the observed estradiol concentration.

Source facts

Action
For domestic customers - the recall is extended to the domestic end user-level. Domestic customers were notified via telephone using a Phone Script (attached in Attachment 5), starting on January 5,2005. Each customer receives a fax-back form and a Technical Bulletin directly after this phone contact. For foreign consignees - the recall is extended to foreign distributors. It is the distributor''s responsibility to extend the recall action to their direct customers. Foreign consignees were notified via e-mail and/or FAX,starting on January 3, 2005.
Initiated
2006-01-03
Root cause
Other
Terminated
2006-08-30
Distribution
Nationwide and Worldwide in Pakistan, Italy, Cyprus, Turkey, Switzerland, Chile, South Africa, Greece, Mexica, Canada, Thialand, Germany, India, United Arab Emirates, Hungary, Israel, Portugal, Taiwan, Austria, Saudi Arabia, Malaysia, Hong Kong, India, Colombia, Argentina, Australia, Costa Rica, Czech Republic, Spain, France, Panama, Brazil, Netherlands, Poland, Sweden, United Kingdom, & El Salvadon.
Recall number
Z-0528-06
Recall status
Terminated
CDRH record ID
43918
FDA classified
2006-02-16
Recall event number
34480

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗