Z-0528-2021: CME America T-Syringe Pump - Neofeed - Product Usage: infusion of medications or fluids requiring continu

Recall record Z-0528-2021 for CME America, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0528-2021
Company in FDA record
CME America, LLC
FDA record creation date
Dec 3, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.

Source facts

Action
On July 30, 2020, the firm sent an "URGENT MEDICAL DEVICE RECALL CORRECTION" Notification Letter via FedEx to customers informing them of two issues: Issue 1: The motor block mechanism may be affected over time by normal wear and tear which may cause the device to stop without alarms being generated. Issue 2: Due to routine cleaning and disinfection, there is a potential for fluid to ingress into the pump and over time, could result in damage to the pump components leading to inoperability of the device. For this issue the device will stop and will alarm. The Recalling Firm is instructing customers to: For Issue 1: Inspect the lead screw to detect the presence of white plastic debris. Inspection must be performed prior to each use of the affected pumps. If signs of wear and tear are observed, discontinue use of the pump and contact the Recalling Firm for product removal. For Issue 2: Follow the Manufacturer Recommended Cleaning instructions in Appendix A and the Technical Service Manual instructions for preventive maintenance provided in Appendix B of the attached Customer Letter to limit the cumulative effects of fluid ingress over time. Immediate Actions to Take: 1) Please refer to the attached customer communication regarding this recall action. 2) As the Recalling Firm plans to notify all affected customers directly, customers are asked to identify customers within their distribution network that purchased the affected product, as defined in the recall notification. Email an excel file listing of all customers to BDRC12@bd.com within 72 hours of receipt of this letter so that the Recalling Firm may initiate customer notification. 3) Complete the attached Distributor Response Form and return to the Recalling Firm's contact noted on the form whether or not the customers have any of the impacted material so that the Recalling Firm may acknowledge the receipt of this notification. 4) Contact the Recalling Firm to arrange for the return of all in sto
Initiated
2020-07-30
Root cause
Other
Terminated
2022-03-08
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of AZ, AK, CA, CO, CT, FL, ID, IL, KS, LA, MA, MI, MO, MT, NC, NE, NV, Oh, OK, OR, PA, TN, TX, WA, and WI. The country of Canada.
Recall number
Z-0528-2021
Recall status
Terminated
CDRH record ID
182567
FDA record created
2020-12-03
Recall event number
86047

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗