Z-0528-2022: MAGEC 1 System

Recall record Z-0528-2022 for NuVasive Specialized Orthopedics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0528-2022
Company in FDA record
NuVasive Specialized Orthopedics, Inc.
FDA record creation date
Jan 26, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Safety Alert to all surgeons informing them of post-implantation fractures of an internal metallic component for device manufactured prior to March 26, 2015.

Source facts

Action
Beginning on 06/25/2019, a Field Safety Notice was communicated to all surgeon informing them that post-implantation fracture of an internal metallic component (i.e., locking pin) has been observed in early versions of the MAGEC System Rods that were manufactured prior to March 26, 2015. Fracture of the locking pin may affect the ability of the device to lengthen and may be associated with Titanium wear debris generation and localized tissue discoloration. Recommended User Actions: - NuVasive, Inc. voluntarily issues this Safety Alert to reinforce the Instructions for Use (IFU) of the MAGEC System to clinical users - The IFU should be consulted on an ongoing basis throughout patient treatment with the MAGEC System. - Users should follow the appropriate postoperative procedure to assess the MAGEC System by X-ray imaging whenever the device is adjusted or at a minimum of once every six months. Typically, the device can be adequately visualized, and a fractured locking pin can be detected, on standard anteroposterior X-ray imaging. Figure 1 shows representative X-ray images for comparison of a device with an intact locking pin (A) to a device manufactured prior to March 26, 2015 with a fractured locking pin (B). A fractured locking pin can typically be identified by examining the noted areas to detect the separation of the internal components and/or the extension of an internal component beyond the device housing. - If a fractured locking pin is detected, removal of the device may be indicated. The decision to remove the device should be made by the physician in consultation with the patient and/or family. - If removal is deemed appropriate by the physician, the device and all associated accessories should be removed and the explanted device should be returned to NuVasive. - Clinical users should review the IFU and be aware of the warnings and precautions on an ongoing basis throughout patient treatment with the device. - Patients and/or families should be reminde
Initiated
2019-06-25
Root cause
Device Design
Distribution
Worldwide
Recall number
Z-0528-2022
Recall status
Open, Classified
CDRH record ID
190711
FDA record created
2022-01-26
Recall event number
89177

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗