Z-0529-2018: Accu2i pMTA Applicator, 14 cm STANDARD, REF/Catalog # 900-600-US, Model # (UPN) H787900600US0, STERILE

Recall record Z-0529-2018 for Angiodynamics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0529-2018
Company in FDA record
Angiodynamics, Inc.
FDA record creation date
Feb 5, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.

Source facts

Action
URGENT VOLUNTARY MEDICAL DEVICE RECALL Letters and Response Forms (dated June 5, 2107) were sent to consignees via FedEx on June 5, 2017. The letter identified the affected product, problem and actions to be taken.
Initiated
2017-06-05
Root cause
Under Investigation by firm
Terminated
2018-09-11
Distribution
Worldwide distribution - US Nationwide
Recall number
Z-0529-2018
Recall status
Terminated
CDRH record ID
156982
FDA record created
2018-02-05
Recall event number
77631

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗