Skip to content

Public FDA record

Z-0530-2008: Terumo Ultrasonic air sensor, 3/8" by 3/32" (9.5 mm x 2.4 mm), Red, for use on Terumo perfusion systems 1

Recall record Z-0530-2008 for Terumo Cardiovascular Systems Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0530-2008
Company in FDA record
Terumo Cardiovascular Systems Corp
FDA classification date
Jan 24, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

False Alarms: The air sensor may malfunction and trigger false alarms, and may continue to alarm thus preventing the device from being reset.

Source facts

Action
Consignees were notified of the defect/issue and what to do if it were to occur (short term work around) via an Urgent Medical Device Removal letter dated 9/14/07. As the permanent fix users were informed the sensor will be replaced when replacements become available.
Initiated
2007-09-19
Root cause
Process control
Terminated
2009-07-07
Distribution
Worldwide: USA, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, Costa Rica, Dominican Republic, Equador, Germany, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, United Arab Emirates, Uruguay, Venezuela and Vietnam.
Recall number
Z-0530-2008
Recall status
Terminated
CDRH record ID
52629
FDA classified
2008-01-24
Recall event number
37989

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗