- FDA identifier
- Z-0530-2015
- Company in FDA record
- Abbott Medical Optics Inc (AMO)
- FDA classification date
- Dec 12, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
AMO has received complaints where the finger grip dislodged while depressing the syringe plunger. The finger grip dislodgement could result in the surgeon's hand unintentionally moving the cannula further than intended, resulting in potential trauma to the eye.
Source facts
- Action
- Abbott Medical Optics issued an Urgent Advisory Notice dated November 4, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised to consult the DFU that is distributed with each device, which contains important warning and precautions for use of the product. Customers were instructed to provide a copy of the Advisory Notice to surgeons and thsoe who need to be aware in their organization. Custommers with questions were instructed to call 877-266-4543. For questions regarding this recall call 714-247-8656.
- Initiated
- 2014-11-06
- Root cause
- Unknown/Undetermined by firm
- Terminated
- 2015-05-19
- Distribution
- Worldwide Distribution - USA (Nationwide and Puerto Rico) and internationally to the following countries; Italy, Austria, Saudi Arabia, Turkey, Hong Kong, Canada, Germany, Australia, Columbia, Venezuela, Netherlands, Portugal, New Zealand, Denmark, Argentina, France, Belgium, Costa Rica, Singapore, Ukraine, Guadeloupe, Kuwait, Chile, Peru, Mexico, Czech Republic, Serbia, Hungary, Slovenia, Guatemala, Finland, Honduras, Malaysia, United Kingdom, Spain, Switzerland, Israel, and Croatia.
- Recall number
- Z-0530-2015
- Recall status
- Terminated
- CDRH record ID
- 131399
- FDA classified
- 2014-12-12
- Recall event number
- 69704