Z-0531-2021: BladderScan Prime Plus Probe REF 0570-0395 - Product Usage: intended to be used for measuring the urine v

Recall record Z-0531-2021 for Verathon, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0531-2021
Company in FDA record
Verathon, Inc.
FDA record creation date
Dec 3, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a change in handle material and process, probe handle may crack.

Source facts

Action
On 10/5/2020, the firm sent a "Field Defect Correction URGENT MEDICAL DEVICE RECALL" Notification to its U.S, Customers via FedEx informing them that due to an unauthorized manufacturing change, the probe handle's plastic may become brittle and crack. Customer are instructed to: 1. Review the information contained within this Recall Notice with all staff members who should be aware of the contents of the letter. 2. Check the list of Serial Numbers that are provided within the Recall Response Form. 3. Perform a visual inspection of the probe handle for cracks or damage. 4. Based on the visual inspection of your device(s), please do one of the following: a. Unboxing and Initial Use " If your BladderScan Prime Plus probe handle plastics have no visible cracks or damage and / or you do not wish for units to be serviced, complete the attached Recall Response Form by indicating no or 0 in the text box and return the form to Verathon by email to CSNotifications@verathon.com. Or " If your BladderScan Prime Plus probe handle plastics have visible cracks and you wish for these units to be serviced, complete attached Recall Response Form by indicating by number the count of probe handle plastics that require repair in the text box and return the form by email to CSNotifications@verathon.com. Verathon Customer Care will contact you to arrange for repair of the probe handle plastics. This replacement will be at no charge to you. b. Subsequent Use If subsequent to submitting your Recall Response Form your BladderScan Prime Plus probes develop cracks (reference Figure 1) during subsequent use and you wish for them to be serviced, please email CSNotifications@verathon.com. Verathon Customer Care will contact you to arrange for repair of the probe handle plastics. This replacement will be at no charge to you. Follow the instructions within the BladderScan Prime Plus Operations and Maintenance Manual (0900-4510 Rev. 11), and as stated below, for conduct
Initiated
2020-09-29
Root cause
Vendor change control
Terminated
2022-06-29
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, NC, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WI, WV and WY. The countries of Austrailia, Belgium, Canada, France, Germany, Kuwait, Luxembourg, New Zealand, Portugal, Singapore, Sweden, Switzerland, Taiwan, and United Kingdom.
Recall number
Z-0531-2021
Recall status
Terminated
CDRH record ID
183915
FDA record created
2020-12-03
Recall event number
86528

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗