If certain keys on the keyboard are pressed during rest ECG procedure, the ECG data may be assigned to the wrong patient.
Source facts
Action
A recall letter dated December 23, 2003 was sent to distributors asking them for contact information for affected customers. Esaote will the provide the user customers with an addendum to the users manual until a software upgrade becomes available and is installed.
Initiated
2003-12-23
Root cause
Other
Terminated
2004-10-28
Distribution
California, Florida, Indiana, Kentucky, Brazil, Colombia and Japan.
Recall number
Z-0532-04
Recall status
Terminated
CDRH record ID
30840
FDA classified
2004-07-20
Recall event number
27980