Skip to content

Public FDA record

Z-0532-2007: Skytron Infinity Series surgical light with triple lightheads, 5 bulbs in each of 3 - 22' diameter lighth

Recall record Z-0532-2007 for Skytron, Div. The KMW Group, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0532-2007
Company in FDA record
Skytron, Div. The KMW Group, Inc
FDA classification date
Feb 22, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

If the mounting plug is improperly adjusted, the unit could separate, come apart, and fall from the mount.

Source facts

Action
The firm initially notified consignees of this problem on 11/23/1999 and again on 11/01/2002 and most lights have already been corrected. Remaining consignees with lights in need of correction were notified of the problem via letter dated 1/8/07 and instructed to contact the distributor to have their lights inspected and modified.
Initiated
2007-01-08
Root cause
Other
Terminated
2009-11-18
Distribution
Nationwide and Canada.
Recall number
Z-0532-2007
Recall status
Terminated
CDRH record ID
50010
FDA classified
2007-02-22
Recall event number
37120

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗