Open original FDA record ↗Recall record
- FDA identifier
- Z-0532-2018
- Company in FDA record
- Spacelabs Healthcare Inc
- FDA classification date
- Jan 25, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Spacelabs has received multiple complaints reporting telemetry beds dropping off the Spacelabs Central Station resulting in a loss of monitoring.
Source facts
- Action
- On January , 2018, a customer letter was sent via priority service, return receipt requested, to all U.S. customers. International consignees will be notified on January 24 , 2018, by email. The letter stated: " Please weigh the benefits versus the risks when deciding whether or not to continue to use your Exhibit Telemetry Receiver(s) until updated as set forth below. We recommend that you brief your staff regarding the failure mode listed above." Spacelabs Healthcare will contact you to schedule a convenient time for Spacelabs to update your Exhibit Telemetry Receiver(s) to software version 1.0.2 at no cost." For further questions, please call (800) 522-7025.
- Initiated
- 2018-01-17
- Root cause
- Under Investigation by firm
- Terminated
- 2018-11-29
- Distribution
- US Distribution to the states of : VA, NJ, MI, GA, and Internationally to France
- Recall number
- Z-0532-2018
- Recall status
- Terminated
- CDRH record ID
- 161031
- FDA classified
- 2018-01-25
- Recall event number
- 78989