Z-0532-2021: CAS-930A Flat Panel Monitor (FPD) used in conjunction with the Canon Interventional Angiography System, M

Recall record Z-0532-2021 for Canon Medical System, USA, INC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0532-2021
Company in FDA record
Canon Medical System, USA, INC.
FDA record creation date
Dec 3, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential that the cable of the Flat panel detector may break during a procedure which may cause the x-ray device to stop before the examination is completed. This could result in a delay with procedure and/or potential damage to the patients blood vessels due to removal without visibility.

Source facts

Action
On 10/08/2020, Canon Medical Systems sent a "Urgent: Medical Device Correction" Notification to all affected consignees via Fax. In addition to informing consignees about the correction, Canon as consignees to take the following actions: 1. On a daily basis, prior to the installation of the new cable with the modified routing, please perform the start-up inspection described in the Operation Manual to confirm that the FPD up and down movement operates normally. Additionally, please have available a mobile device which can perform fluoroscopy such as a surgical C-arm device or another catheter inspection/IVR room. 2. If the above event occurs, please stop using the device and contact your Canon service representative. . 3. Please share this information with all users and reviewing radiologist as well as clinical engineering and the biomedical group at your facility. 4. Please complete and return the attached form and fax it to the toll free number at the top of the form. This form may also be sent via email to RAffairs@us.medical.canon. 5. Thank you for your prompt attention to this matter. If you have any questions, please feel free to contact Terry Schultz Regulatory Affairs Manager at (800) 421-1968 or your local Canon service representative at (800) 521-1968.
Initiated
2020-10-08
Root cause
Component design/selection
Terminated
2022-12-06
Distribution
US Nationwide distribution including in the states of LA, CA, CO, WV, GA, NH, FL, PA, WI, MN, ID.
Recall number
Z-0532-2021
Recall status
Terminated
CDRH record ID
184259
FDA record created
2020-12-03
Recall event number
86650

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗