Z-0532-2022: Vagus Nerve Stimulation (VNS) Therapy Leads, PerenniaDURA, Model: 303. Model 303-20 PereniaDURA Lead (2

Recall record Z-0532-2022 for LivaNova USA, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0532-2022
Company in FDA record
LivaNova USA, Inc.
FDA record creation date
Jan 28, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A manufacturing issue may cause the silicone tubing wall thickness of vagus nerve stimulation leads to be less than the design specification. If the tubing becomes completely breached, the patient may experience discomfort or pain due to delivery of stimulation to an unintended location. If the inner lead conductor is exposed to body fluids the lead may break causing loss of therapy.

Source facts

Action
On 12/02/21, Liva Nova initiated a verbal notification to hospital consignees. Hospitals were asked to remove affected devices from inventory. On 12/17/21, a removal notice was distributed to consignees who were asked to take the following actions: 1. Remove affected devices from your hospital/facility s inventory. 2. Recalling firm will contact you to coordinate removal and replacement of these devices. 4. If affected devices have been implanted, please notify your local representative as soon as possible. A medical device correction notice was sent to doctors managing patients with implanted affected devices were asked to take the following actions: 1. Ask the patient to report any changes in the comfort or perception of their stimulation therapy. 2. Advise patients to use their magnet daily (epilepsy indication only) to verify that stimulation is felt, and to report any change in perceived clinical symptoms related to stimulation. An inability to feel stimulation or changes in clinical symptoms may be an early indicator of a lead break (e.g., increase in seizures, painful stimulation, or changes in the perception of stimulation might indicate a lead break). 3. At each office visit, including those scheduled due to changes reported by patient for items 1 or 2 above, interrogate the device and run device system diagnostics . If diagnostics suggest that a lead fracture is present, program the VNS generator to 0mA Please ensure this notice is communicated to all personnel within your organization who need to be aware of it. This action is being reported to the Food and Drug Administration. For questions customers can contact the recalling firm at: +1 (281) 228-7330 (Monday to Friday, 8 AM to 5 PM CST) or e-mail at cservices@LivaNova.com.
Initiated
2021-12-02
Root cause
Process change control
Terminated
2024-01-25
Distribution
US: NE, MO, AR, WV, KS, MA, IN
Recall number
Z-0532-2022
Recall status
Terminated
CDRH record ID
191291
FDA record created
2022-01-28
Recall event number
89335

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗