Skip to content

Public FDA record

Z-0532-2024: RSP Glenoid Baseplate, 30mm P2 Porous Coated REF 508-32-204. DJO Surgical shoulder device.

Recall record Z-0532-2024 for Encore Medical, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0532-2024
Company in FDA record
Encore Medical, LP
FDA classification date
Dec 12, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Coating hex did not engage with the Straight Mod Hexdriver or with the Hex Head Screw Driver

Source facts

Action
On 9/6/2023, the firm sent an email, "SE190 URGENT PLEASE READ-QUARANTINE NOTICE" stating that "identified a potential issue" with the RSP BASEPLATE, 30MM, W/P2 COATING". This email instructed customers to "Please reply all to this email when you've located the part(s) and have it quarantined. We will provide further instruction as soon as possible. Please do not request an RMA at this time". On 11/10/2023, the firm sent via email an "URGENT MEDICAL DEVICE RECALL" Letter to customers informing them that the RSP Glenoid Baseplate was found to have an undersized hex, preventing engagement with the Straight Mod Hexdriver and /or Hex Head Screw Driver. Customers were instructed to: 1. Pass this notice to all those within their organization or to any organization where the potentially affected product has been transferred. 2. Review their stock for the affected products and lot numbers for the items to be returned. 3. Complete acknowledgement and response form and contact Customer Service at 1-800-456-8696 to place a replacement order and receive an RMA number. The customer will be provided with a shipping label to return the affected product. For questions, contact Livia Kuruvila at (914)450-3931 or by email at livia.kuruvila@enovis.com
Initiated
2023-09-06
Root cause
Component design/selection
Distribution
U.S.: AZ, CA, FL, IN, KS, KY, ME, MN, MS, NJ, OH, Puerto Rico, RI, SC, TN, TX, VA, and WA O.U.S.: None
Recall number
Z-0532-2024
Recall status
Open, Classified
CDRH record ID
204323
FDA classified
2023-12-12
Recall event number
93428

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗