Skip to content

Public FDA record

Z-0532-2026: SJM Pericardial Patch, with EnCap Technology 5x10 cm, Model Number REF C0510; PATCH, PLEDGET AND INTRACA

Recall record Z-0532-2026 for Glycar SA Pty., Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0532-2026
Company in FDA record
Glycar SA Pty., Ltd.
FDA classification date
Nov 25, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The impacted lot may not meet the required tensile strength specification.

Source facts

Action
Glycar issued a Stock Withdrawal Notification to its sole distributor on 10/09/2025. The notice explained the issue and requested the following: "Required Actions 1. Immediately identify and segregate all affected stock in your possession. 2. Clearly label the identified stock as "WITHDRAWN PRODUCT" DO NOT DISTRIBUTE OR USE. 3. Suspend any further distribution or transfer of the affected lots. 4. Provide written confirmation of the quantity and storage location of all affected stock within five (5) business days of receipt of this notification. 5. Our Quality team will coordinate collection, return, or replacement of affected product as appropriate." For questions or require further clarification, contact: Malcolm Applewhite, malcolm@glycar.co.za On 10/17/2025, the firm issued an URGENT: MEDICAL DEVICE RECALL notice the medical facility via email. This notice explained the issue, potential risk to health, and requested the following: "Actions to be Taken by the Customer / User: 1. Identify and quarantine any affected products with lot number T2502257 in your possession. 2. Cease use of the affected device immediately. 3. Share this notice with applicable personnel within your institution. 4. Abbott will coordinate and arrange for return and replacement of affected units. 5. Complete and return the attached Acknowledgment Form within 5 business days to confirm receipt of this notice." For questions or assistance, contact: Regulatory Affairs Manager, Glycar SA (Pty) Ltd, Malcolm Applewhite, malcolm@glycar.co.za
Initiated
2025-10-09
Root cause
Process control
Distribution
US
Recall number
Z-0532-2026
Recall status
Open, Classified
CDRH record ID
216911
FDA classified
2025-11-25
Recall event number
97877

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗