When the error condition occurs the x-ray exposures will terminate immediately. The digital system (EPS-Plus/ADR-1000A) will immediately stop acquiring images or will continue to acquire blank images. The remote console will lock-up upon release of the hand switch. The operator must reboot the system. This will result in the loss of the images acquired during the contrast portion of the study.
Source facts
Action
Recall letter to be sent to to customers of affected devices beginning February 23, 2006. Corrected software to be issued by Toshiba America Medical Systems Inc, for all affected devices.
Initiated
2006-02-23
Root cause
Other
Terminated
2012-02-29
Distribution
Nationwide to AZ, CA, FL, MT, & PA
Recall number
Z-0533-06
Recall status
Terminated
CDRH record ID
45009
FDA classified
2006-03-29
Recall event number
34955
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.