Z-0533-2021: PROCEED Surgical Mesh Hernia Mesh Oval, 15 cm x 20 cm

Recall record Z-0533-2021 for Ethicon, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0533-2021
Company in FDA record
Ethicon, Inc.
FDA record creation date
Dec 3, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm is recalling the PROCEED Surgical Mesh device, because it is a sterile device, and a customer found a hair inside of the primary packaging of a single unit.

Source facts

Action
Note: This recall is specific to this single lot of PROCEED Surgical Mesh. Health care practitioners who have treated patients using this product should follow those patients post-operatively in the usual manner with no additional action required. Product subject to the recall in your inventory can be identified by the product code information described in the above table. all unused PROCEED Surgical Mesh subject to this recall are required to be returned. Please utilize attachment 1 for assistance in identifying the product lots subject to this recall. ACTION REQUIRED: 1. Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). 2. Remove the product subject to this recall and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 3. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 4. Complete the Business Reply Form (Attachment 2) confirming receipt of this notice and return to Ethicon sales representative or distributor by [INSERT DATE]. Please return the Business Reply Form even if you do not have products subject to this recall. 5. Follow instructions in the letter and immediately return unused impacted PROCEED Surgical Mesh subject to this recall to Ethicon sales representative or distributor. To receive credit reimbursement, customers must return product subject to this recall. Any non-affected product and any product returned after the date specified will not receive credit. 6. Keep this notice visibly posted for awareness until all product subject to this recall has been returned to Ethicon Korea. While processing your returns, please maintain a copy of this notice with the product subject to this recall and keep a copy for your re
Initiated
2020-10-23
Root cause
Under Investigation by firm
Terminated
2022-12-21
Distribution
The impacted lot (295 units) was distributed to 27 countries worldwide excluding the US. However, the firm is only recalling the 15 units that were distributed in South Korea due to South Korea's Ministry of Food and Drug Safety (MFDS) requirements.
Recall number
Z-0533-2021
Recall status
Terminated
CDRH record ID
184219
FDA record created
2020-12-03
Recall event number
86631

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗