Z-0534-2008: Ambu, Positive End Expiratory Pressure (PEEP) Valves, PEEP Valve 10, 22/18 mm, REF 000 138 000, Manufact

Recall record Z-0534-2008 for Ambu Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0534-2008
Company in FDA record
Ambu Incorporated
FDA classification date
Dec 22, 2007
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Lower pressure than expected: Medical device for respiratory care may leak and not register accurate flow settings that affect patient respiration.

Source facts

Action
Ambu notified consignees with a letter dated 08/21/07 and flagged as Trouble Shooting Guide. The letter advised that some valves do not hold the set PEEP level under certain conditions and consignees were further advised to consider the guideline as an addendum to existing directions for use. Consignees were instructed to return (to Ambu) for replacement the identified lots listed in the notification and confirm receipt of the advisory by e-m or fax. Ambu subsequently identified additional lots of valves and replacement valve parts for recall and notified all consignees by "Urgent Device Recall" letter dated 11/29/07. The letter advised healthcare practitioners of the impact on patient treatment that may result in less than effective treatment. Consignees were further advised to cease use of the device and return for replacement.
Initiated
2007-08-21
Root cause
Process design
Terminated
2008-06-16
Distribution
Nationwide.
Recall number
Z-0534-2008
Recall status
Terminated
CDRH record ID
66963
FDA classified
2007-12-22
Recall event number
44734

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗