Z-0534-2021: Shimadzu Mobile X-Ray System, MobileDaRt Evolution (MX8 Version), with adjustable handle option, Made in

Recall record Z-0534-2021 for Shimadzu Medical Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0534-2021
Company in FDA record
Shimadzu Medical Systems
FDA record creation date
Dec 4, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential that the adjustable handle option on the X-Ray system may have been installed with improper bolts. This could result in improper driving operation which could lead to harm of individuals around the device.

Source facts

Action
Shimadzu issued a "Urgent: Voluntary Medical Device Recall Notice" on 11/09/2020" via email. In addition to informing consignees about the recalled device, the notification letter asked consignees to take the following actions: 1. If your unit is on the affected list, then please check the drive handle assembly to see if the bolts have loosened. If you feel there is a problem with the mobile drive and movement, please discontinue use of the system and contact your local service provider. 2. Shimadzu has identified the affected units and informed your authorized service representative. They will be contacting you to schedule time to implement the corrective action. If you do not receive a call or visit concerning this matter, please contact Shimadzu National Support. 3. Please complete the "Medical Device Recall Return Response" form and email to drodriguez@shimadzu-usa.com 4. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Initiated
2020-11-09
Root cause
Device Design
Distribution
US Nationwide distribution including in the states of AL, AZ, CA, FL, GA, IA, IL, IN, KY, LA, MI, MN, MS, NC, ND, NE, NJ, NY, OH, OR, PA, SC, TN, TX, WA, WI.
Recall number
Z-0534-2021
Recall status
Open, Classified
CDRH record ID
184407
FDA record created
2020-12-04
Recall event number
86734

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗