Skip to content

Public FDA record

Z-0534-2025: Medline Shoulder Repair Pack-LF, REF DYNJ82607, 2 kits/case, sterile.

Recall record Z-0534-2025 for MEDLINE INDUSTRIES, LP - Northfield. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0534-2025
Company in FDA record
MEDLINE INDUSTRIES, LP - Northfield
FDA classification date
Nov 25, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Some of the suture packages within the convenience kits contain the incorrect needle type/size and suture diameter/length.

Source facts

Action
The recalling firm issued letters on and dated 10/15/2024 via email and first-class mail explaining the issue, potential risk, and informing the consignee stickers are to be provided upon request to be placed onto the Medline pack instructing that the affected component is to be removed and replaced with an item in inventory. Required actions were included which were to (1) Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately; (2) Use the link and information in the letter to complete the response form. List the quantity of affected product in inventory on the form. Upon receipt of the submitted response form, the consignee will receive over-labels to place on affected inventory with instructions for the consignee staff to remove the affected component prior to using the kit. The labels will be provided via FedEx overnight; (4) If the consignee is a distributor or has resold or transferred this product to another company or individual, the consignee is to notify them of the recall communication.
Initiated
2024-10-15
Root cause
Mixed-up of materials/components
Distribution
US Nationwide distribution in the states of AL, GA, MA, NJ, and PA.
Recall number
Z-0534-2025
Recall status
Open, Classified
CDRH record ID
210867
FDA classified
2024-11-25
Recall event number
95638

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.