An issue with the pace detect function was discovered on a 9200 II monitor. Could result in a falsely high or low heart rate reading.
Source facts
Action
On 12/22/04 distribution of new units was placed on hold and customers were notified to recommend that they do not use the pace detect function. On 12/27/03 an email was sent to customers of the solution. On 12/29/04 a Technical Worksheet describing how to replace a resistor at R8 was emailed to customers. Between 12/29/04 and 01/03/05 kits containing instructions and new resistors were sent to customers to begin reworking of the devices.
Initiated
2005-01-05
Root cause
Other
Terminated
2005-11-14
Distribution
Internationally only.
Recall number
Z-0535-05
Recall status
Terminated
CDRH record ID
36761
FDA classified
2005-02-10
Recall event number
30811