Skip to content

Public FDA record

Z-0535-2014: 3M ESPE Imprint II Regular Body , Catalog Number 9379, and Paradigm Regular Body VPS, Catalog Number 5315

Recall record Z-0535-2014 for 3M Company - Health Care Business. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0535-2014
Company in FDA record
3M Company - Health Care Business
FDA classification date
Dec 18, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Specific lots of Imprint II and Paradigm impression material refill packs are being recalled because the moisture displacement feature in both is greatly reduced which may impact the quality of the dental impression.

Source facts

Action
3M ESPE sent an Voluntary Recall Notification Letter dated November 21, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. Customers were instructed to: 1) Review your inventory and identify any products from these lot numbers 2) Quarantine products from these lots and do not further distribute 3) If you have shipped any products from these lots to your customers, inform them of this voluntary recall using the attached Dear Doctor Letter and Inventory Checklist 4) To return and replace product from these lots, contact 3M ESPE Dental Products Customer Service at 1-800-237-1650 If you have any questions, please contact the above number for assistance. Once again, we sincerely apologize for any inconvenience this may cause you.
Initiated
2013-11-21
Root cause
Employee error
Terminated
2014-06-02
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Canada, France, Germany, Hong Kong, Italy, Japan, Korea, Mexico, United Arab and Emirates.
Recall number
Z-0535-2014
Recall status
Terminated
CDRH record ID
124262
FDA classified
2013-12-18
Recall event number
66944

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗