Z-0535-2021: VNS Therapy SENTIVA DUO # 1000-D - Product Usage: indicated for the treatment of chronic or recurrent dep

Recall record Z-0535-2021 for LivaNova USA Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0535-2021
Company in FDA record
LivaNova USA Inc
FDA record creation date
Dec 4, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During internal testing, it was found that upon a device reset, the Generator exhibits the incorrect Model Number when interrogated.

Source facts

Action
On 10/09/2020, the firm hand-delivered an "URGENT: FIELD SAFETY NOTICE" informing its customers that in the event that a generator reset occurs (whether due to troubleshooting efforts or a device issue), the device will erroneously display the incorrect model number as "Model 1000 on the device In-Session programmer screen. The risk to patients could potentially be when the patient undergoes a future generator replacement surgery. Actions Physicians are to take: 1. Model 1000-D devices susceptible to this issue are provided in Attachment 1. When interrogating the generator and performing diagnostic testing per labeling, verify that the In-Session screen displays M1000-D. 2. If an interrogation of the generator is found to display M1000 In-Session or on the Session Report after interrogation, contact Customer Quality at (281) 228-7330 (Monday to Friday, 8 AM to 5 PM CST) or their local sales representative to report the event and for troubleshooting assistance. 3. To complete and return the attached Customer Response Form by fax to (281) 853-1248 or by e-mail to LivaNova.FSCA@livanova.com. The Recalling Firm is developing a device update that will eliminate the possibility for this erroneous model number display and expects that update to be available later in 2020. The Recalling Firm will contact its customers to schedule a visit to update the implanted device. For questions about this Medical Device Correction, contact Customer Quality at +1 (281) 228-7330 (Monday to Friday, 8 AM to 5 PM CST) or e-mail at cservices@LivaNova.com.
Initiated
2020-11-12
Root cause
Device Design
Terminated
2023-01-12
Distribution
International distribution in the countries of Austria and United Kingdom.
Recall number
Z-0535-2021
Recall status
Terminated
CDRH record ID
184499
FDA record created
2020-12-04
Recall event number
86776

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗