Z-0535-2025: VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP

Recall record Z-0535-2025 for Ortho-Clinical Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0535-2025
Company in FDA record
Ortho-Clinical Diagnostics, Inc.
FDA classification date
Nov 25, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

May Generate Negatively Biased Results When Testing Samples with High Concentrations of CRP, showing less than upper AMI of 90 mg/L (in the 70-90 mg/L range) while the samples having higher than 90 mg/L. User no way to know if the sample has to be diluted to measure again.

Source facts

Action
QuidelOrtho issued Urgent Product Correction Notification letter on 10.21.24. Letter states reason for recall, health risk and action to take: " Discontinue using, render unusable, and discard all remaining inventory of lots manufactured from the affected coating(s) of VITROS Chemistry Products CRP Slides. " Complete and return the enclosed Confirmation of Receipt form. Upon receipt of your completed Confirmation of Receipt form, QuidelOrtho will credit or replace your discarded inventory. " Save this notification with your User Documentation or post this notification by each VITROS 250/350/5,1 FS/4600/5600/XT 3400/XT 7600 System in your laboratory until the issue has been resolved. " Please forward this notification if the affected product was distributed outside of your facility. " If your laboratory has experienced the issue with this product and you have not already done so, please report the occurrence to your local Global Services Organization. Root Cause Our investigation is ongoing; however, additional testing has been implemented in our release process to mitigate this issue prior to the release of future coatings of VITROS CRP Slides
Initiated
2024-10-11
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution in the states of AZ, FL, ID, IL, KS, MN, MT, NY, PA, UT and the countries of Australia, Belgium, Bermuda, Brazil, Canada L3R 0B8, Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore 609917, Spain, Sweden, The Netherlands, United Kingdom.
Recall number
Z-0535-2025
Recall status
Open, Classified
CDRH record ID
210845
FDA classified
2024-11-25
Recall event number
95634

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗