Reason stated in the source
Lower pressure than expected: Medical device for respiratory care may leak and not register accurate flow settings that affect patient respiration.
Source facts
- Action
- Ambu notified consignees with a letter dated 08/21/07 and flagged as Trouble Shooting Guide. The letter advised that some valves do not hold the set PEEP level under certain conditions and consignees were further advised to consider the guideline as an addendum to existing directions for use. Consignees were instructed to return (to Ambu) for replacement the identified lots listed in the notification and confirm receipt of the advisory by e-m or fax. Ambu subsequently identified additional lots of valves and replacement valve parts for recall and notified all consignees by "Urgent Device Recall" letter dated 11/29/07. The letter advised healthcare practitioners of the impact on patient treatment that may result in less than effective treatment. Consignees were further advised to cease use of the device and return for replacement.
- Initiated
- 2007-08-21
- Root cause
- Process design
- Terminated
- 2008-06-16
- Distribution
- Nationwide.
- Recall number
- Z-0536-2008
- Recall status
- Terminated
- CDRH record ID
- 66965
- FDA classified
- 2007-12-22
- Recall event number
- 44734