Skip to content

Public FDA record

Z-0536-2013: Biomet Polaris 5.5 Button Lock Screw Inserter. Catalog number 14-500178

Recall record Z-0536-2013 for Ebi, Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0536-2013
Company in FDA record
Ebi, Llc
FDA classification date
Dec 17, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

Broken screw insertor hardness was measured and found to be below the acceptable limits.

Source facts

Action
Biomet mailed Urgent Medical Device Recall Notice/Fax Back Response Form dated 9/25/2012 to the distributors on 9/29/2012.
Initiated
2012-09-29
Root cause
Process control
Terminated
2015-12-07
Distribution
Nationwide Distribution, including the states of MO, TX, and IN.
Recall number
Z-0536-2013
Recall status
Terminated
CDRH record ID
114420
FDA classified
2012-12-17
Recall event number
63653

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗