Z-0536-2021: 3PLUS1 Multilevel Plasma Calibrator Set, Order No. 75128 - Product Usage: The kit is intended for amino a

Recall record Z-0536-2021 for Chromsystems Instruments & Chemicals GmbH. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0536-2021
Company in FDA record
Chromsystems Instruments & Chemicals GmbH
FDA record creation date
Dec 4, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Products do not meet stability criteria and may impact results.

Source facts

Action
Consignees are informed of this issue by Field Safety Notice FSN 03-2020 on 11/02/2020, also addressing the measures that should be taken by the consignees: If the products listed above were already used for measurements, consignees shall check with the attending physician(s) whether it is necessary to review the results of asparagine, homocystine and methionine. Consignees shall stop using the reconstituted products listed above that are stored more than 6 weeks at < -18 C for measurements of asparagine, homocystine and methionine. Consignees shall destroy the remaining stock of reconstituted calibrators and controls that are stored more than 6 weeks at < -18 C and were intended to be used in measurements of asparagine, homocystine and methionine according to their valid laboratory regulations. Chromsystems will replace stocks. Calibrators and controls will now have a reduced in-use stability (after reconstitution) for asparagine, homocystine and methionine of 6 weeks at < -18 C. For all other analytes, in-use stability is not changed. If consignees have forwarded any of the products listed above to another laboratory, consignees shall inform that laboratory of the content of this letter and forward a copy. Consignees shall document their measures on provided Response document (attachment Response) and return their response until 20.11.2020 to Chromsystems (contact details provided in the attachment Response). Consignees shall make sure that all users of the products mentioned above and other persons affected in their organization are informed about the FSN. If they have provided third parties with these products, they shall forward a copy of this information or inform the contact person of Chromsystems.
Initiated
2020-11-02
Root cause
Under Investigation by firm
Terminated
2021-12-20
Distribution
Worldwide distribution - US Nationwide distribution in the states of CA and NY; The countries of Australia, Belgium, Brazil, Germany, Finland, France, Greece, Israel, Italy, Canada, Croatia, Kuwait, Austria, Russia, Switzerland, South Africa, Taiwan, Turkey, and United Arab Emirates.
Recall number
Z-0536-2021
Recall status
Terminated
CDRH record ID
184386
FDA record created
2020-12-04
Recall event number
86719

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗