Open original FDA record ↗Recall record
- FDA identifier
- Z-0537-2010
- Company in FDA record
- Bock,Otto,Orthopedic Ind,Inc
- FDA classification date
- Dec 16, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Otto Bock Health Care LP has discovered that in some cases, the securing screws of the joint axes may not have been properly mounted. This can cause the axes to gradually move out of the joint and result in a risk of falling for the user.
Source facts
- Action
- Consignees were sent on 10/28/08 a Otto Bock "Urgent: Correction Notice" letter dated October 27, 2008. The letter described the product and problem. It provided instructions if the product was in stock and instructed the consignee to contact the patients (a patient letter was provided) and to complete and return the reply card. Consignees were to contact Otto Bock at their Customer Service to request a return authorization.
- Initiated
- 2008-10-28
- Root cause
- Employee error
- Terminated
- 2011-12-10
- Distribution
- The product was only distributed within the US by the US company. Any OUS distribution was handled by the manufacturer in Germany.
- Recall number
- Z-0537-2010
- Recall status
- Terminated
- CDRH record ID
- 75488
- FDA classified
- 2009-12-16
- Recall event number
- 50486