Potential failure to audibly alarm due to speaker hardware component failure.
Source facts
Action
The firm sent a letter dated December 31, 2003 instructing their customers to install an audio cable from the CPU to the Flat Panel Monitor to provide redundant speaker capability. The audio cable and installation instructions were included in the letter. The letter also included a Correction Response Form which customers should fill out upon completion of correction.
Initiated
2003-12-31
Root cause
Other
Terminated
2005-01-06
Distribution
The firm distributes the product nationwide and internationally.
Recall number
Z-0538-04
Recall status
Terminated
CDRH record ID
30993
FDA classified
2004-02-13
Recall event number
28003