Open original FDA record ↗Recall record
- FDA identifier
- Z-0538-2011
- Company in FDA record
- Elekta, Inc.
- FDA classification date
- Dec 3, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The HeadFIX Baseplate is not screwed down to the adapter to allow to compensate for roll and pitch with the HeadFIX leveling screws of the HeadFIX Baseplate.
Source facts
- Action
- Elekta, Inc. sent an Important Notice A343 dated October 14, 2010 to all affected consignees. The Notice identified the product, the problem, and the action to be taken by the consignees. Consignees were instructed constantly monitor the patient before and during the treatment. Check the patient's position after every patient movement (e.g. coughing). Advise all appropriate personnel of the content of the notice place a copy of the Notice In the Important Notice Section of the HeadFix User Manual P10103-108. For questions regarding this recall service.schwabmuenchen@elekta.com
- Initiated
- 2010-10-14
- Root cause
- Other
- Terminated
- 2010-12-28
- Distribution
- Worldwide Distribution - USA including AZ, CA, CT, FL, GA, IL, KY, MI, MS, NY, NC, OK, OR, RI, TN, TX, VA, WA and the country of Canada
- Recall number
- Z-0538-2011
- Recall status
- Terminated
- CDRH record ID
- 95410
- FDA classified
- 2010-12-03
- Recall event number
- 57111