Skip to content

Public FDA record

Z-0539-2010: Holding Forceps for Trident Acetabular System; non Sterile; Catalog number 2105-0000; Howmedica Osteonics

Recall record Z-0539-2010 for Stryker Howmedica Osteonics Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0539-2010
Company in FDA record
Stryker Howmedica Osteonics Corp.
FDA classification date
Dec 16, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The tips of the forceps were larger than the positioning holes located on the trials and the implant liners.

Source facts

Action
Important Product Correction letters, dated January 24, 2007, were sent to all consignees on via Federal Express. The letter stated the problem and asked customers to examine their inventory and hospital locations to identify the affected product. All affected forceps product should be retrieved from hospitals and returned to their branch or agency warehouse for reconciliation. Customers are also to reconcile all affected product on the attached Product Accountability Form and fax a copy of that form to 201-831-6069 within 5 days of receipt of the letter. All affected forceps product is to be destroyed and discarded at their branch location. Questions should be directed to their Stryker Orthopaedics Customer Service rep at 1-866-OR-ASSIST.
Initiated
2007-01-24
Root cause
Pending
Terminated
2009-12-21
Distribution
Worldwide Distribution -- Including countries of Hong Kong, Australia, Japan, New Zealand, Hong Kong, Singapore, Germany, Sweden, The Netherlands, France, Spain, Portugal, and the United Kingdom.
Recall number
Z-0539-2010
Recall status
Terminated
CDRH record ID
82817
FDA classified
2009-12-16
Recall event number
52384

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗