Open original FDA record ↗Recall record
- FDA identifier
- Z-0540-2020
- Company in FDA record
- GE Healthcare, LLC
- FDA classification date
- Dec 3, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.
Source facts
- Action
- The firm disseminated its notices by letter beginning on 10/11/2019. The notices identified the following safety issues: - The bedside panels of Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation can be upright and look closed but not be latched. - The portholes also can look closed when not latched. - If a canopy cover is used, it can hold the bedside panel or porthole door closed without being latched. The product may continue to be used following specific safety instructions outlined in the notice. On/about 11/8/2019, the firm issued an "Amended - Urgent Medical Device Recall" letter that supplemented the previous notification to provide additional safety instructions and materials (poster regarding risk of patient fall, user manual addendum, and labeling, including safety labels). Update 6/3/2021: In the event a new consignee was identified, GE Healthcare has sent the original letter.
- Initiated
- 2019-10-11
- Root cause
- Device Design
- Terminated
- 2024-08-15
- Distribution
- worldwide
- Recall number
- Z-0540-2020
- Recall status
- Terminated
- CDRH record ID
- 177149
- FDA classified
- 2019-12-03
- Recall event number
- 84077