Skip to content

Public FDA record

Z-0541-2021: Clinical Innovations ROM Plus Collection Kit IFU: ROM Plus Rupture of Membrane Test ROM-5025, ROM-6025

Recall record Z-0541-2021 for Clinical Innovations, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0541-2021
Company in FDA record
Clinical Innovations, LLC
FDA record creation date
Dec 7, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Possible false negative or false positive results. The incorrect sample collection swabs (non-scored) were assembled into assembled into the in vitro diagnostic kits. If the user does not break the swab stick correctly, or if scissors or other tools are used to cut the swab handle false negative or false positive results can occur.

Source facts

Action
On August 17, 2020 through August 19, 2020, the firm sent an "URGENT SAFETY NOTICE" to customers via email informing them that the incorrect swabs (non-scored) were package with their in vitro diagnostic kits and that new instructions and training material will be provided. Customers are instructed to identify any affected products and to complete the response card that is attached to the Customer Notification Letter. Customers are also provided the option of returning the affected products to the Recalling Firm. A company representative will contact and arrange for the return of the product For any questions contact the Recalling Firm at +1-801-268-8200.
Initiated
2020-08-13
Root cause
Process change control
Terminated
2024-05-24
Distribution
US: CO, IL, IN, MO, OH, RI, UT, and VA. OUS: None
Recall number
Z-0541-2021
Recall status
Terminated
CDRH record ID
183211
FDA record created
2020-12-07
Recall event number
86255

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗